To ensure accurate readings, perform a zero and span calibration during commissioning. Let the sensor stabilize for at least 30 minutes after power-up before starting this procedure.

Gently pull the display module forward and unclip its retaining tether to expose the terminal blocks underneath. Power and 4-20mA Signal Wiring (3-Wire Current Source) Xnx Universal Transmitter Quick Start Guide

XNX Universal Transmitter Quick Start Guide: Installation, Wiring, and Calibration To ensure accurate readings, perform a zero and

The default passcode to access calibration and configuration menus is 0000 . Calibration Procedure Power and 4-20mA Signal Wiring (3-Wire Current Source)

: The XNX cover must be fully seated with a minimum of 9 threads engaged to maintain its explosion-proof rating.

This is where most technicians get stuck. The Xnx terminal block is powerful but crowded. You will likely use a 3-wire or 4-wire shielded cable (18–22 AWG).