Narishige Pc10 | Manual New _top_
Clean the glass capillaries with ethanol or acetone prior to loading to remove dust and manufacturing oils. Contaminants on the glass cause uneven heating and asymmetric tips. Step 2: Loading the Sample Loosen the upper and lower clamping knobs.
A heavier weight (using 4x weights instead of 2x) results in a shorter, wider pull.
The secondary heating phase triggers automatically, followed immediately by the final drop that breaks the glass into two identical pipettes. Carefully remove both pipettes from the clamps. 5. Parameter Optimization and Troubleshooting
Later, when a new student asked where to find the PC-10 manual, Lena would hand them the box and say, simply: “Read it. Then teach it to your hands.” narishige pc10 manual new
Ensure the power outlet is properly grounded and matches the voltage specified on the device’s rear serial plate (typically 100V, 120V, or 220-240V).
If you have a , please specify — I can write a separate guide for those digital models.
Narishige PC-10 Glass Pipette Puller: Setup, Operation, and Maintenance Manual Clean the glass capillaries with ethanol or acetone
If the heater coil becomes brittle or fails to heat (Error E_99), it must be replaced. Environment:
Clean the guide rails with a lint-free cloth and apply a microscopic drop of high-grade machine oil. Preventive Maintenance
The heater coil becomes extremely hot during operation. Never touch the coil while the unit is powered on. A heavier weight (using 4x weights instead of
Use a damp cloth to clean the rod and ensure the movement is smooth. Do not use oils or lubricants on the main shaft. Conclusion
Switches operation between single-stage (Step 1 only) and dual-stage (Step 1 and Step 2 sequentially) modes.
Are you experiencing any specific with your new unit? Share public link
Operation centers on balancing heat and pulling force to create the desired pipette. This is controlled by two primary parameters: